Chapter 3 Preparing a Run
3-8 Applied Biosystems 7900HT Fast Real-Time PCR System and SDS Enterprise Database User Guide
Quick Review: Powering On the 7900HT Instrument
Powering On the
Instrument
Note: The following table includes a set of abridged procedures for activating the
components of the Applied Biosystems 7900HT Fast Real-Time PCR System. For a
complete explanation of the procedure, see “Powering On the 7900HT Instrument”
on page 2-4.
IMPORTANT! Do not power on the 7900HT instrument while the lower-side panel of
the instrument is removed. Doing so prevents the instrument from uploading the
firmware from the computer and causes the software to display an error.
1. Power on the monitor and computer.
2. If using an automation module, power on the Zymark
®
Twister Microplate
Handler.
3. Power on the 7900HT instrument.
4. Start the SDS software (select > Programs >
Applied Biosystems > SDS 2.2.1 > SDS 2.2.1)
5. If using the SDS Enterprise Database, enter your user name and password in the
appropriate fields of the Login dialog box, then click .
The SDS software displays the program window.
HI-POT
AB CD
GR1728
Power switch
(Rear panel of the Twister)
GR2396
7900HT
Power button