Step 5 – Programming the Plate Document Method
Applied Biosystems 7900HT Fast Real-Time PCR System and SDS Enterprise Database User Guide 3-19
Step 5 – Programming the Plate Document Method
User Access
Requirement
If using the SDS Enterprise Database, you must belong to the Scientist or
Administrator User Group to modify the method of a plate document.
About SDS
Methods
During a run, the SDS software controls the instrument based on the instructions
encoded in the method of the plate document. Each new plate document
(except allelic discrimination) contains a default method that you must customize for
your experiment.
Methods contain the:
• Thermal Cycler Conditions
• Auto Increment Values
•Ramp Rates
• Data Collection Options
• Reaction Volume Setting
Database
Considerations
Configure the method of your plate document to collect data only in the necessary
stages of the PCR. Because of memory constraints, the SDS Enterprise Database
cannot save real-time plate documents greater than 40 MB. The file size for a
real-time run depends on the length of the run and the configuration and placement
of the data collection icons. Longer runs and runs configured to collect data at
multiple stages of the method can be considerably larger than the 15–25 MB average.
To create a method for… Then…
• absolute quantification
• relative quantification
program the method as explained on page 3-21.
IMPORTANT! If using the SDS Enterprise Database, read
the discussion of considerations below.
allelic discrimination see page 3-20.
dissociation curve analysis To perform a dissociation curve analysis do the following:
1. Program the method for the absolute quantification
experiment as explained on page 3-21.
2. Add a temperature ramp to the method for dissociation
curve analysis as explained on page 3-23.

Table of Contents

Other manuals for Applied Biosystems 7900HT

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Applied Biosystems 7900HT and is the answer not in the manual?

Applied Biosystems 7900HT Specifications

General IconGeneral
BrandApplied Biosystems
Model7900HT
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals