Chapter 4 Operating the Instrument
4-6 Applied Biosystems 7900HT Fast Real-Time PCR System and SDS Enterprise Database User Guide
Preventing Contamination
Contamination
and the
Nuclease
Assay
PCR using the 5´ nuclease assay requires special laboratory practices to avoid false
positive amplifications (Kwok and Higuchi, 1989). The assay’s logarithmic
amplifications can potentially amplify a single DNA molecule (Saiki et al., 1985;
Mullis and Faloona, 1987).
General PCR
Practices
Observe the following guidelines when assembling optical plates or loading
TaqMan
®
Low Density Arrays for use on the 7900HT instrument:
Wear a clean lab coat (not previously worn while handling amplified PCR products
or used during sample preparation) and clean gloves when preparing samples for
PCR amplification.
Change gloves whenever you suspect they are contaminated.
Maintain separate areas, dedicated equipment, and supplies for:
Sample preparation
PCR setup
PCR amplification
Analysis of PCR products
Never bring amplified PCR products into the PCR setup area.
Open and close all sample tubes carefully. Try not to splash or spray PCR samples.
Keep reactions and components capped as much as possible.
Use positive-displacement pipets or aerosol-resistant pipet tips.
Clean lab benches and equipment with 10% bleach solution.
Compatible Consumables
Consumables for
Use with
the 7900HT
Instrument
Table 4-1 Compatible Consumables
Consumable Type
Compatible Seals
Use these… Not these…
ABI P
RISM
384-Well Optical
Reaction Plates
•ABIPRISM
Optical Adhesive Covers
(quantity 100)
Optical Adhesive Covers (quantity 25)
384-Well Septa