DRAFT
September 1, 2004 11:39 am, CH_Basic.fm
Chapter 4 Operating the Instrument
4-16 Applied Biosystems 7900HT Fast Real-Time PCR System and SDS Enterprise Database User Guide
Centrifuging the TaqMan Low Density Arrays
Equipment and
Materials Needed
You need the following equipment and materials to centrifuge the Low Density
Arrays:
TaqMan Low Density Array(s), containing sample-specific PCR reaction mix
(page 4-14)
•Sorvall
®
Legend T Centrifuge
•Sorvall
®
/Heraeus Custom Buckets (four required), each with its own card holder
Blank balance cards
Placing Cards in
the Buckets
Note: The centrifuge holds four Sorvall/Heraeus buckets. Each bucket holds up to
three Low Density Arrays (loaded and/or blank balance cards).
1. Obtain an empty Sorvall/Heraeus Custom Bucket and a card holder for the
centrifuge.
IMPORTANT! The Sorvall/Heraeus buckets and card holders required for the
Low Density Arrays are custom-made. Do not use any other bucket/card holder
system for this procedure.
2. Place the bucket on a lab bench, with the label facing you.
3. Insert Low Density Arrays into the card holder, making sure that:
The fill reservoirs project upwards out of the card holder
The reaction wells face the same direction as the “This Side Out” label
IMPORTANT! Use blank balance cards to fill any remaining positions in the
racks.
Note: The card holder supports the Low Density Array fill reservoirs during
centrifugation.
4. Place a filled card holder in the bucket so that the “This Side Out” label faces
the front of the bucket, which has the Sorvall emblem on it.
GR2155
GR2160
Sorvall
emblem here

Table of Contents

Other manuals for Applied Biosystems 7900HT

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Applied Biosystems 7900HT and is the answer not in the manual?

Applied Biosystems 7900HT Specifications

General IconGeneral
BrandApplied Biosystems
Model7900HT
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals