Preparing TaqMan Low Density Arrays for Use
Applied Biosystems 7900HT Fast Real-Time PCR System and SDS Enterprise Database User Guide 4-15
Loading the Fill
Reservoirs with
Sample-Specific
PCR Reaction
Mix
1. When the packaging has reached room temperature and you are ready to load
PCR reaction mix, carefully remove a Low Density Array from its packaging.
2. Place the Low Density Array on a lab bench, with the foil side down.
3. Load 100 µL of the desired sample-specific PCR reaction mix into a 100-µL
micropipette.
4. Hold the micropipette in an angled position and place the tip in the fill port.
Note: There is a fill port on the left arm of each fill reservoir; it is the larger of
the two holes.
IMPORTANT! Do not allow the tip to contact and possibly damage the coated
foil beneath the fill port.
5. Dispense the sample-specific PCR reaction mix so that it sweeps in and around
the fill reservoir toward the vent port.
IMPORTANT! Pipette the entire 100 µL into the fill reservoir.
IMPORTANT! Do not allow the tip to contact and possibly damage the coated
foil beneath the fill port.
IMPORTANT! Be careful when pushing the micropipette plunger to its second
stop position (to expel the sample-specific PCR reaction mix from the tip). If a
large amount of air is released, it can push the reaction mix out of the fill
reservoir via the vent port.
6. To transfer the sample-specific PCR reaction mix from the fill reservoirs into
the reaction wells, the Low Density Array must be centrifuged.
Continue with “Centrifuging the TaqMan Low Density Arrays” on page 4-16.
GR2158
Fill port
Vent port
GR2159

Table of Contents

Other manuals for Applied Biosystems 7900HT

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Applied Biosystems 7900HT and is the answer not in the manual?

Applied Biosystems 7900HT Specifications

General IconGeneral
BrandApplied Biosystems
Model7900HT
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals