DRAFT
September 1, 2004 11:39 am, CH_Real-Time.fm
Chapter 6 Analyzing Real-Time Data
6-50 Applied Biosystems 7900HT Fast Real-Time PCR System and SDS Enterprise Database User Guide
• If analyzing a Relative Quantification Multiple Plate Document, view the
results of the run displayed within the Gene Expression plot and
Results table.
Eliminating Wells
from the Analysis
The SDS software allows you to omit wells from the analysis. If you have identified
outliers or data you want to exclude from the analysis, follow the procedure below to
remove the corresponding wells from use.
IMPORTANT! If a well is removed from use before the plate document is run, the
SDS software does not collect data for the well during the run. However, if a well is
removed from use after the plate document has been run, the software excludes the
well from the analysis but does not delete the data.
1. If omitting wells from a Relative Quantification Multiple Plate Document,
select a plate in the Plates field that contains wells you want to omit.
2. While pressing and holding the Ctrl key, select the well(s) in the plate grid that
you want to omit.
3. If omitting wells from a Relative Quantification Multiple Plate Document,
repeat steps 1 through 2 for each plate in the study.
4. Select the Setup tab.
5. In the well inspector, click the Omit Well(s) check box to exclude the selected
wells from the analysis.
Note: To reinstate an omitted well, select the well, then click the Omit Well(s)
check box.
Omit Well(s) check box
(selected)
Selected wells
(removed from use)

Table of Contents

Other manuals for Applied Biosystems 7900HT

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Applied Biosystems 7900HT and is the answer not in the manual?

Applied Biosystems 7900HT Specifications

General IconGeneral
BrandApplied Biosystems
Model7900HT
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals