13.1 Requirements for the Transport
13.2 Requirements for Operation
13.2.1 Setting Up and Fastening the Modules
The modules must be positioned securely and, for example, fastened in the vehi
-
cle:
- Position the module or the module stack on a smooth, level underground.
- The module or the module stack must always be securely fastened.
- The module or the module stack may be sloped by no more than 15°.
13.2.2 Heat Exchange with the Environment
When operating the modules, an efficient heat exchange with the environment
must be ensured:
- Keep the module away from heat sources and protect the module against
direct exposure to the sun.
- Maintain minimum distances from the module to the top and a distance of at
least 5 cm on both longitudinal sides of the system.
- The free space above and behind the module must be selected so that suf-
ficient air circulation is ensured.
- Do not cover the ventilation slots of the module!
- Do not place the module or the module stack on one of the longitudinal sides
with ventilation slots.
- Secure the module or module stack in such a way that ventilation is not com-
promised.
- Do not operate the module in completely enclosed containers.
NOTE
Possible damage or destruction of electronics!
Transport the modules of the system only with protective flap closed and pro-
tective cap attached!
WARNING
Risk of injury due to inadequate fastening
- Secure the product so that it does not move uncontrollably.
- Only use carrier systems and fastening materials that can accommodate the
static and dynamic forces of the product and are suitable for the ambient
conditions.