MFG-2000 Series User Manual 
366 
-221 Settings conflict; infinite burst changed trigger source to MANual 
Example: The trigger source is changed to 
immediate from manual when infinite burst mode is 
selected. 
-221 Settings conflict; burst period increased to fit entire burst 
Example: The function generator automatically 
increases the burst period to allow for the burst 
count or frequency.   
-221 Settings conflict; burst count reduced 
Example: The burst count is reduced to allow for the 
waveform frequency if the burst period is at it’s 
maximum. 
-221 Settings conflict; trigger delay reduced to fit entire burst 
Example: The trigger delay is reduced to allow the 
current period and burst count. 
-221 Settings conflict;triggered burst not available for noise 
Example: Triggered burst cannot be used with noise. 
-221 Settings conflict;amplitude units changed to Vpp due to high-Z load 
Example: If a high impedance load is used, dBm 
units cannot be used. The units are automatically set 
to Vpp. 
-221 Settings conflict;trigger output disabled by trigger external 
Example: The trigger output terminal is disabled 
when an external trigger source is selected. 
-221 Settings conflict;trigger output connector used by FSK 
Example: The trigger output terminal cannot be used 
in FSK mode. 
-221 Settings conflict;trigger output connector used by burst gate 
Example: The trigger output terminal cannot be used 
in gated burst mode.