PEL-2000A Series User Manual
102
Local loads
Local mode operation is useful to quickly test loads using the load
module control panel rather than the mainframe control panel.
Local load modules can be configured to operate independently to
the mainframe. This can be useful when settings need to remain
unchanged on the mainframe. Note however that the local modules
cannot change the modes (CC, CV, CR, CP), only the values.
Group Unit mode is not supported for local module control.
DUT
Step
Description
Details
1. Setup
Ensure the channel load is set up
as desired.
Pages 38, 45
2. Channel
selection
Ensure the correct channel or
Value (A/B) is selected by using
the R/L or A/B key.
Page 120
3. Measurement
Mode
selection
If in CC or CR mode, Static or
Dynamic mode can be selected.
CC Pages 121,
132
CR Pages 136,
139
4. Run the Load
Press the LOAD key to start/end
loading the device under test.
Page 122
Optional
5. Short
configuration
Configure the SHORT settings.
Page 189
6. Display
To change the display output, use
the DISPLAY key.
Page 123
7. Shorting the
load
To short the load, use the SHORT
key.
Page 123