Run Program
Background
The Program function on the PEL-2000A series
supports a total of 12 different programs at any
one time with 10 sequences to each program. Up to
12 programs can be chained together. The Program
function is able to create a number of Go/NoGo
tests.
Run Program is not supported in Group Mode
(page 72).
Program
Sequence
A program sequence is simply a single load test. A
program is a battery of each of these tests run in
succession. Each sequence loads the settings for
each channel from Memory Data (Memory MXXX).
The Memory Data stores settings such as the
operating mode and range for each channel. Each
sequence loads all channels at the same time,
unless programmed otherwise. Sequences for each
channel run synchronously.
Time
CH1
Sequence
001
Sequence
002
Sequence
003
CH2
CH3
CH4
CH5 CH5 Active = Off
Each Sequence has a number of configuration
options that apply to all the channels equally.
Sequence Item
Description
Memory
Loads the channel settings for
each load module
Range: M001~M120