Run Program   
 
The Program function on the PEL-2000A series 
supports a total of 12 different programs at any 
one time with 10 sequences to each program. Up to 
12 programs can be chained together. The Program 
function is able to create a number of Go/NoGo 
tests.  
Run Program is not supported in Group Mode 
(page 72). 
A program sequence is simply a single load test. A 
program is a battery of each of these tests run in 
succession. Each sequence loads the settings for 
each channel from Memory Data (Memory MXXX). 
The Memory Data stores settings such as the 
operating mode and range for each channel. Each 
sequence loads all channels at the same time, 
unless programmed otherwise. Sequences for each 
channel run synchronously.  
Time
CH1
Sequence 
001
Sequence 
002
Sequence 
003
CH2
CH3
CH4
CH5 CH5 Active =  Off
Each Sequence has a number of configuration 
options that apply to all the channels equally. 
Loads the channel settings for 
each load module 
Range: M001~M120