OPERATION
193
3. Press the Selector knob to edit
the selected setting, turn to
change the setting from Panel to
External.
CH CONT External
4. Press the Selector knob or Enter
to confirm selection.
OR
ENTER
Channel control is now activated.
To turn Channel Control off, CH
CONT must be set to Panel again.
When Channel Control is active,
EXT will be displayed on the side
panel for the active channel.
CH1
conf
CCDH
80V
EXT
Note
Channel Control can only be activated on the active
channel; other channels will not be affected.
Configuring the Independent Setting
The Independent setting allows a channel to be
load independent from the mainframe. What this
means is that a load module with Independent set
to ON can only load from the local load module. If
the LOAD ON/OFF key is pressed from the
mainframe, the channel with Independent set to
ON will be unaffected by the mainframe, except
when running a program.
Independent
ON/ OFF