PEL-2000A Series User Manual
106
Parallel Load Modules
The Group Unit setting allows for a quick and easy parallel setup
for load modules of the same type and rating. However, the Group
Unit settings only apply for CC and CR modes.
DUT
Step
Description
Details
1. Setup
Choose the appropriate load
modules and make sure they are
installed. All load modules must
be of the same type and rating.
Page 38
2. Connection
Connect the terminals to the DUT.
Page 45
3. Group Unit
mode
configuration
Enable Group Unit Mode and
configure.
Page 204
4. Measurement
mode
selection
Select measurement mode (CC,
CR).
CC Page 128
CR Page 134
5. Range
selection
Set the range to high or low (CC,
CR mode).
CC Page 128
CR Page 135
6. Mode
selection
Choose Static or Dynamic mode.
CC Pages
129, 132
CR Pages 136,
139
7. Dynamic
levels
For Dynamic mode, set the
dynamic levels, slew rate and
timers.
CC Page 130
CR Page 136
8. Static Values
For Static mode, set the A(B)
Value and slew rate.
CC Page 132
CR Page 139