1. Press the CHAN button and use
the selector knob to select a
channel.
2. Press F1 until CP mode is
displayed in the display panel.
ConfigureConfigure
Mode
CP
Range
Low
Changing the operating mode will only affect the
current (active) channel. Other channels will not be
affected by any changes.
Editing CP Parameters
Constant Power mode can be set to have a
maximum limit (Curr Limit). Using the current
limit enables limiting the current draw.
When using CP mode on a single channel load
module, two power levels can be configured, A
Value and B Value. On a dual channel load
module, only one power level can be configured
per channel: A Value.
Ensure the menu is in CP Mode. See page 141.
ConfigureConfigure
Mode
CP
Range
Low