OPERATION 
141 
4.  Repeat steps 1-3 for B Value (if applicable), 
rising and falling SlewRate. 
The last Value (A Value or B Value) that is set 
becomes the active setting. To swap between A Value 
and B Value, use the A/B keys on the local load 
module (not applicable to Group Unit mode). 
For Group Unit mode, use the Switch Value 
parameter to switch between A and B Value. 
A/B Value and rising/falling SlewRate can be set for 
both High and Low Range.  
*Press Shift to toggle between coarse and fine 
adjustment when editing the A Value and B Value 
parameters. See page 197 for details. 
Select CV Mode 
 
The PEL-2000A series electronic load operates in 
four different modes, Constant Current (CC), 
Constant Resistance (CR), Constant Voltage (CV), 
and Constant Power (CP).  
CV Mode cannot be used with the Group Unit 
mode. 
When a channel is active, the F1 
key can be used to switch 
between each operating mode. 
1.  Press the CHAN button and use 
the selector knob to select a 
channel. 
2.  Press F1 until CV mode is 
displayed in the display panel.