The rising and falling slew rate (A/uS) determines 
the speed at which the load levels change from A 
to B Value (Static mode) or from Level1 to 2 
(Dynamic mode) and vice versa. 
Go/NoGo is also supported. Center, High and 
Low limits can be set as either percentages or 
voltage values. A delay time of up to 1 second can 
also be set. 
In Constant Voltage Mode the load units will sink 
current whilst keeping the voltage constant.  
Single channel load modules support 2 values (A 
Value, B Value) and have an adjustable cut-off 
current limit. Dual channel load modules only 
have A value. 
Response speed can also be set to fast (Fast) or 
slow (Slow). The response speed relates to the 
slew rate of the current response.   
Go/NoGo functionality is also supported either as 
a percentage or as a current value.