OPERATION
167
3. Press the Selector knob to edit
Repeat, then turn to increase or
decrease the value.
OR
4. Use the number pad to enter a
number.
Select 0 to choose infinity.
CAL.
987
3
CLEAR
21
0
4 5 6
P0 LOCK
P3P1 P2
P6P4 P5
P9P7 P8
Repeat Times0005
5. Press the Selector knob or Enter
to confirm selection.
OR
ENTER
6. Repeat for the remaining parameters.
7. Press Save (F3) to save the loop.
F3
Channel Duration Time Settings
Each sequence can have the timing duration data
of another sequence. For example CH1’s sequence
can import the timing duration settings of CH2’s
sequence.
This is useful to quickly compare two different
loads to the same timing characteristics. See page
77 for more details.
Each channel’s sequence can be turned off by
configuring the channel Setting to OFF. If a
channel uses the same channel number, i.e. CH 01
01, then the duration time settings will not be
altered for that channel.