OPERATION
197
Configuring Step Resolution
The CC, CR, CV and CP step resolution settings
can be edited in the configuration menu. These
step resolution settings directly correspond to the
step resolution of the coarse adjustment when
setting the CC, CR, CV and CP parameters.
The minimum and maximum step resolution that
can be set for each channel is dependent on the
load module. For more information on step
resolution see page 86.
Minimum*
1
Maximum*
2
Unit
CCH Step
HR/20000
HR/2
Amperes A
CCL Step
LR/20000
LR/2
Amperes A
CRH Step
HR/40000
HR/2
Siemens
CRL Step
LR/40000
LR/2
Siemens
CVH Step
HR/40000
HR/2
Voltage V
CVL Step
LR/40000
LR/2
Voltage V
CPH Step
HR/10000
HR/2
Watt W
CPL Step
LR/10000
LR/2
Watt W
CCH Step
HR/40000
HR/2
Amperes A
CRH Step
HR/40000
HR/2
Siemens
CRL Step
LR/40000
LR/2
Siemens
CVH Step
HR/40000
HR/2
Voltage V
CVL Step
LR/40000
LR/2
Voltage V
CPH Step
HR/30000
HR/2
Watt W