OPERATION
181
Setting (OCP/OVP/OPP/UVP)
Over Protection is used to set the voltage, current
or power limit. In the event that the current,
voltage or power exceeds the over protection
settings, the load module display will show an
error message and beep an alarm.
When tripped, Under Voltage Protection (UVP)
will turn off the load. UVP trips when the load
voltage drops below a set limit.
Only when the protection settings are set to On
(XXP Setting -On) will the protection modes be
active.
All protection settings can be set to 2% higher
than specification rating.
OCP Level
1.25% Rating A ~ 102% Rating A
OCP Setting
ON/OFF/Clear
OVP Level
1.25% Rating V ~ 102% Rating V
(0.5% Rating V ~ 102% Rating V
for PEL-2041A)
OVP Setting
ON/OFF/Clear
OPP Level
PEL-2020A : 1W ~ 102W
PEL-2030A(L) : 0.9W ~ 30.6W
PEL-2030A(R) : 1.25W ~ 255W
PEL-2040A : 1.75W ~ 357W
PEL-2041A : 1.75W ~ 357W
OPP Setting
ON/OFF/Clear
UVP Level
OFF ~ current using operating
voltage range of slave module.
UVP Setting
Clear
Protection Clear
All