OPERATION
135
ConfigureConfigure
Mode
CR
Dynamic
Range
Low
Note
Changing the operating mode will only affect the
current (active) channel. Other channels will not be
affected by any changes.
Select CR Range
Constant Resistance mode can run in high and
low range. Range is dependent on the load
module.
Ensure the menu is in CR Mode. See page 134.
ConfigureConfigure
Mode
CR
Dynamic
Range
Low
1. Press the F2 (Range) key repeatedly
until High or Low range is selected.
Range
High
Range
Low
F2
The range will be reflected in both the
bottom menu system and the Current
Operation Channel Status panel.
CRL
CH1
main
80V
CR Static Low
CR Static High
CR Dynamic Low
CR Dynamic High
CRL
CRH
CRDL
CRDH