INTERFACE
289
Input: active low (0-1V)
active high (4-5V)
Note: Input type is internally pulled up to 5V
with a 10k resistor.
Output: high (floating)
low (0-1V)
Note: Output type is internally Open collector
outputs, maximum 30VDC with 1.1V saturation
voltage (100mA).
When Enable (pin9) is on (active low), the
following is disabled from the mainframe: Load
On/Off (pin 10) activating loads and recalling
preset (pin 1-3) or setup memory (pin 5-8).
Maximum 5 (1 master + 4 slave units) devices
can be linked altogether with a maximum cable
length of 30cm for each cable.
All the connected devices must be turned on.
No loop or parallel connections
Configure Go/NoGo Interface
Use a DSUB (DB-15 Female) connector
to connect to the Go/NoGo port.
The Go/NoGo port is an output only
port.
GO / NG OUTPUT
1 8