OPERATION
163
Edit Sequence
The Sequence function can be configured to create
a unique load profile to accurately simulate loads
in real time for single or multiple loads. Sequence
can only be used with CC static or CR static
modes. For details see page 77.
Each Sequence is composed of a number of points
with customizable current/resistance, slew rate
and duration times. Each sequence can be looped
an infinite amount of times. Sequences are only
applicable for CC (Static) and CR (Static) modes.
Note
The sequence function should not be confused with
program sequences. They are not the same. Program
sequences cannot be used with the Sequence
function and vice versa.
Value
Setting Min ~ Setting Max /A
SlewRate
Load module dependent
SlewRate
Duration Time
0.000025 ~ 60,000 seconds
1. Choose a channel and mode.
Page 126,
128
2. Press the CHAN key, F4 (Seq.
Edit) to enter the Sequence Edit
menu.
CHAN
F4