Add Point will insert a new point directly after the
current point. The value of the current/resistance of
the inserted point will be the average of the point
before and after. All other settings will remain
unchanged.
9. To delete the current point, press
Delete Point (F2).
10. Use the selector knob to change
the current point number.
The Point number can only be changed if more points
have already been added.
11. Press Save (F3) to save the
sequence.
The save icon will only appear after a change has
been made in the menu.
Create Sequence Loop
Sequences can be looped a number of times. The
loops can be started at any point in the sequence.
The Start of Loop function determines which
point will start each repeating loop.
The On End of Seq. function will hold the load
current (of the selected sequence) to a designated
value until all the other sequences have finished
running.
CC Vrange sets the range in CC mode for
Sequences. See page 77 for more details.