130
Note
Changing from static to dynamic mode will only affect
the current (active) channel.
Editing CC Dynamic Parameters
Dynamic Constant Current Mode has two
operating current levels, slew rates and timers.
Slew rates determine the speed at which the load
will change from one level to the next.
The timers determine how long the load
module/channel will stay at level 1 or level 2.
Ensure the menu is in CC Dynamic Mode. See
page 129.
ConfigureConfigure
Mode
CC
Dynamic
Range
High
Level1
0 ~ Setting Max A
Level2
0 ~ Setting Max A
SlewRate
Load module dependent
SlewRate
Timer1
0.025 ~ 30000.0 ms
Timer2
0.025 ~ 30000.0 ms
Note
When used in Group Unit mode, the Level1 & Level2
range is the combined rating of all the units used in
Group Unit mode.
1. Use the Selector knob to
highlight Level1.