PEL-2000A Series User Manual
136
Note
Changing the range will only affect the current
(active) channel. Other channels will not be affected
by any changes.
All resistance values and slew rates are dependent on
Range, i.e., A Value in low range can be different from
A Value in high range.
Select CR Dynamic Mode
Constant Resistance mode can be set to dynamic
or static mode. Dynamic mode is used to
automatically set varying load rates.
Ensure the menu is in CR Mode. See page 134.
ConfigureConfigure
Mode
CR
Dynamic
Range
Low
1. Press F3 until Dynamic Range
mode is selected.
Dynamic
Static
F3
Note
Changing from static to dynamic mode will only affect
the current (active) channel.
Editing CR Dynamic Parameters
Dynamic Constant Resistance Mode has two
operating resistance levels, slew rates and timers.
Slew rates determine the speed at which the load
will change from one level to the next.