PEL-2000A Series User Manual
134
4. Repeat steps 1-3 for the
remaining parameters.
Note
The last Value (A Value or B Value) that is set
becomes the active setting. To swap between A Value
and B Value, use the A/B keys on the local load
module. This is not applicable to Group Unit mode.
For Group Unit mode, use the Switch Value
parameter to switch between A and B Value.
A/B Value and rising/falling SlewRate can be set for
both High and Low Range.
*Press Shift to toggle between coarse and fine
adjustment when editing the A Value and B Value
parameters. See page 197 for details.
Set to CR Mode
The PEL-2000A series load operates in four
different modes, Constant Current (CC), Constant
Voltage (CV), Constant Resistance (CR) and
Constant Power (CP). Constant Resistance mode
will maintain a constant resistive load, using
variable current and voltage levels.
When a channel is active, the F1
key can be used to switch
between each operating mode.
F1
Mode
CC
Mode
CR
Mode
CV
Mode
CP
1. Press the CHAN button and use
the selector knob to select a
channel.
LOCAL
CHAN
2. Press F1 until CR mode is
displayed in the display panel.
F1