OPERATING DESCRIPTION
73
CCDH
ConfigureConfigure
Mode
CC
Dynamic
Range
High
Level2
SlewRate
SlewRate
Timer1
Timer2
Level1 90
10
0.80
0.80
0.025
0.025
A
A
A/uS
A/uS
mS
mS
main
80V
CH1P
LOAD
USB
CCDH
ConfigureConfigure
Mode
CC
Dynamic
Range
High
Level2
SlewRate
SlewRate
Timer1
Timer2
Level1 30
10
0.80
0.80
0.025
0.025
A
A
A/uS
A/uS
mS
mS
main
80V
CH1S
LOAD
USB
When used in CR mode, the equation for
equivalent resistance for all the parallel loads is:
For example, if 2 load modules have a set
resistance of 100Ω each, the equivalent resistance
of the load modules would be 50Ω. The Level
setting in Para mode would be 50Ω and 100Ω in
Sync mode.
CRDL
ConfigureConfigure
Mode
CR
Dynamic
CH1P
Range
High
Level2
SlewRate
SlewRate
Timer1
Timer2
Level1 50.0000
100.000
0.40
0.40
0.025
0.025
Ω
Ω
A/uS
A/uS
mS
mS
main
80V
LOAD
USB
CRDL
ConfigureConfigure
Mode
CR
Dynamic
CH1S
Range
High
Level2
SlewRate
SlewRate
Timer1
Timer2
Level1 100.000
100.000
0.40
0.40
0.025
0.025
Ω
Ω
A/uS
A/uS
mS
mS
main
80V
LOAD
USB
Display mode determines what units are displayed
on the local load modules: V/I, V/W, I/W, S. The
displayed units can only be controlled through this
menu.