OPERATION
143
CVH
ConfigureConfigure
Mode
CV
CH1
B Value
Curr Limit
A Value 10.00
15.00
10.00
V
V
A
Slow
Response
Slow
main
80V
Range
High
06/15/18
16 : 50
2. Press the Selector knob to edit
the selected value, then turn to
increase or decrease the value*.
OR
Use the number pad to enter a
number.
CAL.
987
3
CLEAR
21
0
4 5 6
P0 LOCK
P3P1 P2
P6P4 P5
P9P7 P8
A Value V10.00
3. Press the selector knob or Enter
to confirm selection.
OR
ENTER
4. Repeat steps 1-3 for the remaining parameters.
Note
The last Value (A Value or B Value) that is set
becomes the active setting. To swap between A Value
and B Value, use the A/B keys on the local load
module.
When setting the current limit, please ensure that the
current limit is within the test device’s limits.
A/B Value can be set for both High and Low Range.
*Press Shift to toggle between coarse and fine
adjustment when editing the A Value and B Value
parameters. See page 197 for details.