PEL-2000A Series User Manual
178
12. Press Save (F3) to save the OCP
test automation parameters.
F3
13. Press the Load key to start to
OCP Test Automation.
ON /
OFF
LOAD
Configure
Run OCP Test
OCP
USB
06/15/18
16 : 50
CH: 1
4.8550
V
4.8550V
1.10A
Voltage Reading: The voltage of the DUT before
the OCP was triggered.
Current Reading: The current of the DUT before
the OCP was triggered.
Note
In addition to the setting the OCP test parameters as
described above, the VON voltage settings must also
be set according to the output characteristics of the
DUT.