OPERATION
183
5. Use the Selector knob to
highlight OCP Setting.
6. Use the selector knob to turn
ON, OFF or CLEAR the OCP
Setting.
7. Repeat steps 1-5 for :
OCP Level
OVP Level
OVP Setting
OPP Level
OPP Setting
UVP Level
UVP Setting
When any of the protection settings are tripped,
Alarm will be shown on the Mainframe Status
Panel and an alarm tone will sound by default.
Alarm
On the local load module, the protection setting
that has been tripped will be displayed.
OCP
OVP
REV*
OPP
OTP*
CPP*
UVP
8. Turn the load off by pressing the
Load key and turn off the load
input.
ON /
OFF
LOAD