OPERATING DESCRIPTION
69
CV + CC Mode
Load Input
Voltage
Input Current
Current Limit
Response Speed
Response speed can be set to fast or slow. Fast
response and slow response is determined by the
load module specifications. Slow response speeds
are suitable for large loads as quick current
changes will induce induction which can cause
large voltage drops. The PEL-2000A series will try
to rectify any voltage drops. However if voltage
drops are too large, they may cause the load to go
into oscillation. Large voltage drops caused by line
voltage induction may damage the machine.
Range
Fast
Slow
1kHz
100Hz
Go/NoGo
Go/NoGo testing can be with either current
(Ampere) values (High, Low) or percentage values
(Center, High %, Low %). A delay time of up to 1
second can also be set.
CV Mode: GO/NOGO
Voltage
Load
Current
High
Low
Load
Voltage
Go
NoGo
NoGo