PEL-2000A Series User Manual
92
Voltage Monitor
The input voltage, like the load input current can
be externally monitored with the channel control
connectors. The VMON pin of the channel control
connector outputs a voltage of 0~10V to represent
the load input voltage as a percentage (0~100%) of
the rating voltage.
Voltage Monitor
Voltage
Monitor
Output
Load Input
Voltage
10V
0V
Rating voltage0V
To determine the Voltage Monitor Output
(VMON) , use the following formula;
V
VRating
voltageinputLoad
VMON 10
Where “Rating V” is the rating voltage of the load
module.
Turning on the
Load
A load is turned on when Load On input is set to
On (active low). A load is turned off when the
Load On input is set to Off (active high).
When a load is turned on from the channel control
interface, the load can be turned off from the
mainframe, local module and via remote control.
However the opposite is not true; when a load is
turned off using the channel control interface, the
load cannot be turned on via the mainframe, local
module or via remote control.