3-10 624074/07
3 Hardware options
3.1.3.3 Module exhaust
The exhaust port on the rear of the module contains barbs for
attaching scavenging tubing. The port contains two different
sized barbs, small and large.
The small barb is sized for 0.060 inch (1.524 mm) ID and up to
0.100 inch (2.54 mm) OD.
The large barb is sized for 0.125 inch (3.175 mm) ID and up to
0.375 inch (9.525 mm) OD.
3.1.3.4 Connecting the CO
2
sidestream sensor
WARNING
• Always ensure the integrity of the patient
breathing circuit after insertion of the sample
kit by verifying a proper CO
2
waveform (capno-
gram) on the ventilator display.
• If the capnogram appears abnormal, calibrate
the CO
2
sensor, inspect the CO
2
sample kit and
replace if needed.
• Monitor the capnogram for higher-than
expected CO
2
levels during ventilation. These
can be caused by sensor or patient problems.
• Do not use the CO
2
sensor module if it appears
to have been damaged or if it fails to operate
properly. Refer servicing of the module to HAM-
ILTON MEDICAL authorized personnel.
• To reduce the risk of explosion, DO NOT place
the CO
2
sensor in a combustible or explosive
environment (for example, around flammable
anesthetics or other ignition sources).
• DO NOT operate the CO
2
sensor when it is wet
or has exterior condensation.