624074/07 3-39
Figure 3-28. User Interface for cuff controller, showing
full deflation of cuff
3.5.7 Alarms
If during automatic or manual cuff-pressure adjustment a leak-
ing cuff is detected, the ventilator alarms with a low-priority
alarm. After pressing
Deflate
, a window asking the user to
confirm the deflation appears. This helps avoid unintentional
deflation and loss of pressure by accidentally pressing
Deflate
. Disconnection of the connecting tubing between
tracheal tube and ventilator results in a medium-priority dis-
connection alarm (see section 10.5).
3.6 Working with the integrated HAMILTON-H900
humidifier
NOTE:
• Integration with the ventilator system is available
only with the HAMILTON-H900 humidifier. You can
use other supported humidifiers with the system,
without the integration.
• The integrated humidifier is not available in all
markets. Check with your distributor to determine
availability.