A-20 624074/07
A Specifications
A.6 Monitored parameters
Table A-8 is an alphabetical list of monitored parameter ranges
and resolutions, including the Vent Status panel.
Table A-9 lists the ranges of the real-time curves. Pressure,
flow, and volume measurements are based on readings from
the Flow Sensor. Optionally, you can select an auxiliary pressure
sensing site via the Paux connector. Directly measured (non-
calculated) parameters have these tolerances:
• Volume-related parameters: ±10% or 2 ml, whichever is
greater (ATPD) for calibrated infant/neonatal flow sensor;
±15% or 20 ml, whichever is greater (ATPD) for calibrated
adult/pediatric flow sensor.
• Pressure-related parameters: ±5% of reading or 1 cmH
2
O,
whichever is greater
• Oxygen: ±3% absolute FiO
2
, including stability of measure-
ment
•CO
2
(BTPS):
0 to 40 mmHg (0 to 5.3 kPa): ±2 mmHg (0.3 kPa)
41 to 70 mmHg (5.4 to 9.3 kPa): ±5% of reading
71 to 100 mmHg (9.4 to 13.3 kPa): ±8% of reading
101 to 150 mmHg (13.4 to 20.0 kPa): ±10% of reading
SpO
2
Range:70 – 80% ±3%
80 – 100% ±2%
Pulse: 30 – 300 b/min ±3%, ±1 b/min