624074/07 11-11
To disinfect the device parts:
1. Clean, but DO NOT re-assemble.
2. Disinfect with a mild bactericidal chemical solution. Accept-
able chemicals include: Schülke & Mayr Lysetola AF and
Gigasepta FF, Henkel-Ecolab Incidura, Sekusepta PLUS, and
CIDEX. Solutions such as these have been tested according
to the manufacturers’ guidelines. Other brand names with
similar active ingredients may also be suitable.
3. Reassemble and reinstall parts, and perform any required
tests.
11.2.3 General guidelines for autoclave sterilization
To Autoclave the device’s parts:
1. Clean.
2. Reassemble.
3. Inspect.
4. Autoclave.
5. Perform any required tests.
11.3 Preventive maintenance
NOTE:
• HAMILTON MEDICAL recommends that you docu-
ment all maintenance procedures.
• Dispose of all parts removed from the device accord-
ing to your institution’s protocol. Comply with all
local, state, and federal regulations with respect to
environmental protection, especially when disposing
of the electronic device or parts of it (for example,
oxygen cell, batteries).
• HAMILTON MEDICAL recommends you to perform
preventive maintenance. No maintenance causes loss
of warranty.