EasyManua.ls Logo

Hamilton G5 - P;V Tool Pro Window

Hamilton G5
594 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
9-10 624074/07
9 P/V tool maneuver option
Figure 9-4. P/V Tool Pro window
1
Reference.
Pushing the reference button overlays up to
3 curves starting at the current displayed PV Tool curve.
The most recent curve is displayed on the top.
2 Plot number/Total number of stored plots
3 Lets you scroll through the previous stored PV Tool curves.
If
History
is active, lets you move the cursor between the
overlaid curves.
4 Settings
1
. Shows the currently active P/V Tool Pro control
settings. These settings are:
Pstart.
PEEP to be applied at start of maneuver.
Ptop.
Maximum pressure to be applied.
End PEEP.
PEEP to be applied at end of deflation limb.
If Pstart and End PEEP are different. A window for con-
firmation displays asking whether you want to change
PEEP/CPAP setting after maneuver.
Ramp speed.
Rate of change of pressure.
Tpause.
Length of pause between inflation and
deflation limbs.
•Push the
Settings
button to open a window which
allows to change the settings.

Table of Contents