Figure 3-23. Selecting the cuff pressure controller
3.5.5 Operation of the automatic cuff
pressure controller
WARNING
Cuff pressure will only be generated and adjusted
if the operator has activated the cuff pressure
adjustments in the System -> IntelliCuff window.
CAUTION
Use regularly the inflating tube, if available on the
tracheal tube and tracheostomy tube to check
proper operation of the automatic cuff pressure
controller.
The device is designed to be immediately used without any
calibrations.
1. First, connect the disposable tubing on the ventilator.
2. Then connect the open end of the disposable tubing to the
inflating tube of the tracheal tube or tracheostomy tube.