1.3.3 Ventilation unit
Figure 1-10 and Figure 1-11 show the patient breathing circuit
connections and other important parts of the ventilation unit.
Figure 1-10. Front of ventilation unit (patient breathing
circuit connections)
19 Main monitoring parameters (MMP). Five numeric parame-
ters set during configuration.
20 Alarm limits. Upper and lower alarm limits, where applicable,
for each MMP.
1 Paux connector. Connects to an auxiliary pressure
sensing site, for use as the pressure input in addition to
the proximal flow sensor measurement. By default there
is no rinse flow through the Paux connector.
2 Pneumatic NEBULIZER connector
3 FLOW SENSOR connection. Always attach the blue
tube to the blue connector and the clear tube to the
silver connector. The blue tube should always be toward
the patient.