I-2 624074/07
I Communication interface
I.1 Introduction
WARNING
To reduce the risk of excessive leakage current due
to ground loops, and to prevent electromagnetic
interference (EMI), ensure the connecting cable has
a high-quality shield and is grounded properly on
one side only, either at the ventilator or receiving
device.
NOTE:
Additional equipment connected to medical electri-
cal equipment must comply with the respective IEC
or ISO standards (e.g. IEC 60950 for data processing
equipment). Furthermore all configurations shall
comply with the requirements for medical electrical
systems (see IEC 60601-1-1 or clause 16 of the 3Ed.
of IEC 60601-1, respectively). Anybody connecting
additional equipment to medical electrical equip-
ment configures a medical system and is therefore
responsible that the system complies with the
requirements for medical electrical systems. Atten-
tion is drawn to the fact that local laws take priority
over the above mentioned requirements. If in doubt,
consult your local representative or the technical
service department.
This interface includes EMI-protective covers for the
connectors. When the connectors are not in use,
ensure the covers are installed.
Ensure the option is enabled before you use the
interface (see Section J.9).
Check ...
messages, which indicate a setting con-
flict, are not sent to the remote alarm outlet. To
show all low-priority alarms, HAMILTON MEDICAL
recommends you read them directly from the
device’s touch screen or consult the event log.