EasyManua.ls Logo

Hamilton G5 - Installing the Pneumatic Nebulizer

Hamilton G5
594 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
624074/07 3-15
3.3 Installing the pneumatic nebulizer
WARNING
Do not use an expiratory filter continuously or
an HME in the patient’s breathing circuit during
nebulization. Nebulization can cause an expira-
tory side filter to clog, substantially increasing
flow resistance and impairing ventilation.
To prevent the expiratory valve from sticking
due to nebulized medications, use only medica-
tions approved for nebulization and regularly
check and clean or replace the expiratory valve
membrane.
Connect the nebulizer in the inspiratory limb
per your institution’s policy and procedures.
Connecting the nebulizer between the flow
sensor and the tracheal tube increases dead
space ventilation.
Be aware that nebulization affects delivered
oxygen concentration.
NOTE:
In the configuration mode you can view and change
the nebulizer settings.
Nebulization is disabled during heliox administration
and in the infant/neonatal application.
The nebulization feature provides a stable driving pressure to
power a pneumatic nebulizer connected to the nebulizer out-
let, optimally specified for 6 to 7 l/min flow. Be aware that
nebulization affects delivered oxygen concentration.
Connect the nebulizer and accessories as shown in Figure 3-9.
Table 1-1 has information about compatible nebulizers.

Table of Contents

Related product manuals