EasyManua.ls Logo

Hamilton G5 - Limits: Setting Alarm Limits

Hamilton G5
594 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
5-28 624074/07
5 Ventilator settings
5.7.1 Limits: Setting alarm limits
WARNING
Ensure the alarm limits are appropriately set
before you place the patient on the device pre-
venting possible patient injury.
Although you can set all alarms rapidly using
the
Auto
alarm function, some settings are not
appropriate under all clinical conditions. HAM-
ILTON MEDICAL recommends: you set the
alarms manually when possible. If circum-
stances require to use the
Auto
alarm function,
verify the correctness of the settings at the
earliest opportunity.
NOTE:
Always ensure the high pressure alarm is correctly
set. This alarm provides a safety pressure limit for the
device to appropriately adjust the inspiratory pres-
sure necessary to achieve the target tidal volume.
The maximum available inspiratory pressure for
INTELLiVENT
®
-ASV, ASV
®
, APVcmv and APVsimv is
10 cmH2O below the high pressure limit, indicated
by a blue colored band on the pressure waveform
display. Set the high pressure limit to a safe value
(example, 45 cmH2O, this limits the pressure target
to a maximum of 35 cmH2O). When the high pres-
sure alarm is set too low, it can lack enough margin
for the device to adjust its inspiratory pressure to
deliver the target tidal volume.
In the ASV
®
and the INTELLiVENT
®
-ASV controls
window, the high pressure alarm will be changed by
adjusting the Pasv limit (pressure target 10 cmH2O
below high pressure alarm limit).

Table of Contents

Related product manuals