EasyManua.ls Logo

Hamilton G5 - Required Conditions for Use; Indication for Use

Hamilton G5
594 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
624074/07 9-3
9.1.2 Required conditions for use
Make sure the following conditions are met before attempting
a P/V Tool maneuver.
The patient must be intubated and ventilated but not
breathing spontaneously. Some patients can require heavy
sedation or neuromuscular blockade to prevent spontane-
ous breathing efforts.
There must be no gas leak throughout the entire system
composed of the ventilator, the breathing circuit, and the
ventilated patient.
Nebulization must be deactivated.
The flow sensor must perform optimally.
Total respiratory system compliance must be a minimum of
5 ml/cmH2O. (The P/V Tool maneuver does not function
correctly with the IngMar infant lung model.)
The P/V Tool is inactive in SPONT, NIV, NIV-ST, nCPAP-PS and
in Backup modes.
The P/V Tool is inactive
Flow sensor calibration
needed
Alarm is active.
9.1.3 Indication for use
As long as the required conditions are met and no contraindi-
cations exist, the P/V Tool maneuver may be performed on any
patient.
The P/V Tool maneuver is particularly valuable for use with
restrictive lung “stiff lung” diseases (for example, ALI or ARDS
patients).
Be careful when performing the P/V Tool maneuver on patients
with obstructive “soft lung” diseases (for example, COPD). Set
Ptop
low to prevent generation of excessive volumes.

Table of Contents

Related product manuals