EasyManua.ls Logo

Hamilton G5 - Fire and Other Hazards

Hamilton G5
594 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
624074/07 ix
tor of the ventilator must still maintain full responsibility for
proper ventilation and patient safety in all situations.
Do not silence the audible alarm when leaving the patient
unattended.
Do not use the exhaust port of the expiratory valve for
spirometry. Due to the HAMILTON-G5’s base flow, the
exhaust gas output is larger than the patient’s actual
exhaled volume.
Heliox gas is not compatible with INTELLiVENT
®
-ASV use.
The Masimo Rainbow SpHb sensor is valid for a total of 66
hours of operation. Once this time expires, you must
connect a new SpHb sensor. For details, see Appendix F,
Pulse oximetry.
Fire and other hazards
To reduce the risk of fire or explosion, do not place the
HAMILTON-G5 in a combustible or explosive environment
(for example, around flammable anesthetics or other
ignition sources). Do not use it with any equipment
contaminated with oil or grease.
To minimize the risk of fire, do not use high-pressure gas
hoses that are worn or contaminated with combustible
materials like grease or oil.
To reduce the risk of fire, use only breathing circuits
intended for use in oxygen-enriched environments. Do not
use antistatic or electrically conductive tubing.
In case of fire, immediately secure the patient’s ventilatory
needs, switch off the HAMILTON-G5, and disconnect it
from its gas and electrical sources.
Do not put a vessel filled with a liquid on the ventilator.
If a liquid enters the product, a fire and/or electric shock
may occur.

Table of Contents

Related product manuals