EasyManua.ls Logo

Hamilton G5 - Spo 2 Alarm Limits Window; Table of Pulse Oximeter-Related Alarm Settings; Adjustable Alarms with Automatic and Standard Settings

Hamilton G5
594 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
624074/07 F-27
F. 6.2 Sp O
2
Alarm Limits window
The Pulse rate alarm and Masimo Rainbow SET parameters
alarm limit controls are displayed in the
Alarms
-
>
Limits
2
window.
Pulse
rate
, is always displayed on the left side,
regardless of which sensing options are installed.
Figure F-5. SpO
2
Alarm Limits 2 window
For details on setting alarm limits, see Section 5.7.1.
F.6.3 Table of pulse oximeter-related alarm settings
Be sure all desired alarms are enabled (Section F. 8. 1 .5 ) and
correctly set. For troubleshooting, see Table F-12.
Table F-7. Adjustable alarms with automatic
and standard settings
Parameter Unit Range
For all settings, -- indicates the alarm is off.
Pulse rate l/min Low: 30 - 230 in increments
of 5 (default: 50)
High: 35 - 235 in increments
of 5 (default: 140)
SpO
2
(Nihon-Kohden)
% Low: 70 - 99 (default: 90)
High: 71 - 100 (default: 99)

Table of Contents

Related product manuals