J-2 624074/07
J Configuration
J.1 Introduction
During configuration, you set up the ventilator with a default
language, breath timing control philosophy, main monitoring
parameter display, and nebulization parameters.
Operator configuration also lets you enable and disable alarms
and set up the I:E timing outlet. You typically configure the
ventilator when you first acquire it, before you put it in service
on a patient.
NOTE:
For details on INTELLiVENT
®
-ASV and Quick Wean con-
figuration, see the INTELLiVENT
®
-ASV manual.
For details on SpO
2
configuration, see Appendix F, Pulse
oximetry.
J.2 Begin configuration
1. To begin the Configuration process, put the ventilator into
Standby.
2. Press the and buttons at the same time.
The
Configuration
button is then displayed.
3. Touch the
Configuration button
. The Configuration
window opens (Figure J-1).
4. Begin device configuration, identify your requirements, and
then activate them.
For more detail on specific instructions, read the next sub-
sections to configure your ventilator.
5. Touch
Close
when done with each panel.
6. When you are done configuring the ventilator, touch
Close/Save
to save your selections.
7. Return to the Patient setup window.