624074/07 A-33
A.8 Configuration specifications
Table A-11 lists the configurable parameters, ranges, and
default or standard (factory) settings.
Rate low b/min 0-158 12 - Volume
modes/ASV:
Rate/fControl-
40%
- All other
modes:
fTotal-40%
Apnea time s 15-60 15 -- 5
SpO
2
high % Off,
71-100
95 Same as stan-
dard settings
1
1. “Off” only for NIV and NIV-ST, low priority alarm “Low ExpMinVol alarm off” is appearing
when set to “Off”
2. “Off” only for NIV and NIV-ST, low priority alarm “Low ExpMinVol alarm off” is appearing
when set to “Off”
Table A-10. Adjustable alarms with automatic and standard settings
Parameter Unit Range Standard
setting
Automatic
setting
Resolu-
tion
Table A-11. Configuration specifications
Parameter Range Standard
setting
Resolu-
tion
Customize
Breath timing philos-
ophy (Controls)
I:E/Pause, Peak Flow/Tip,
%Ti/Pause
I:E/Pause --
Active alarms
(Alarms)
ExpMinVol high,
Pressure low, Vt high,
Vt low, Rate high, Rate
low, Apnea time, VLeak
1
ExpMinVol
high, Pressure
low, Vt high,
Vt low, Rate
high, Rate low,
Apnea time
--