624074/07 F-49
F.11 Masimo specifications
The specifications for the Masimo pulse oximeter with
Rainbow SET technology are provided in Table F-13.
Table F-13. Masimo pulse oximeter specifications
Feature Specifications
Display ranges Oxygen saturation (SpO2): 0 - 100%
Pulse rate (bpm): 25 - 240 bpm
Carboxyhaemoglobin saturation (SpCO): 0 - 99%
Methaemoglobin saturation (SpMet): 0 - 99.9%
Total haemoglobin (SpHb): 0 - 25 g/dL
Respiratory rate (RRa): 0 - 70 breaths per minute
Total oxygen concentration (SpOC): 0 - 35 ml/dl
Hematocrit (SpHct): 0 - 75%
Perfusion Index (PI=): 0.02 - 20%
Pleth Variability Index (PVI): 0 - 100%
Accuracy
See footnotes 1, 2, 3, 4, 5, 6, 7
SpO2, no motion 60 - 80 ± 3% adults/pediatrics/infants
70 - 100% ± 2% adults/pediatrics/infants;
± 3% neonates
SpO
2
, motion 70 - 100% ± 3%, adults/pediatrics/infants/neonates
SpO
2
, low perfusion 70 - 100% ± 2%, adults/pediatrics/infants/neonates
Pulse rate, no motion 25 - 240 ± 3 bpm, adults/pediatrics/infants/neonates
Pulse rate, motion 25 - 240 ± 5 bpm, adults/pediatrics/infants/neonates
Pulse rate, low
perfusion
25 - 240 ± 5 bpm, adults/pediatrics/infants/neonates
SpCO 1 - 40 ± 3%, adults/pediatrics/infants
SpMet 1 - 15 ± 1%, adults/pediatrics/infants/neonates
SpHb 8 - 17 ± 1 g/dl (arterial or venous), adults/pediatrics