EasyManua.ls Logo

Hamilton G5 - Alarms; Masimo Spo 2 Alarm Delay; F.6 Alarms

Hamilton G5
594 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
F-26 624074/07
F Pulse oximetry
F.6 Alarms
The SpO
2
pulse oximeter has several related alarms. You enable
them in the
Configuration
window (Section F.8.1.4) and set
limits in the
Alarms
-
>
Limits
2
tab. You work with them in
the same manner as other ventilator alarms. For alarm settings,
see Table F-7. For troubleshooting, see Table F-12.
NOTE:
If you are running the SpHb sensor in Spot Check mode,
the SpHb low or SpHb high alarms are only active for 1
minute.
F.6.1 Masimo SpO
2
alarm delay
Because SpO
2
can be relatively variable and potentially gener-
ate a high number of alarms, Masimo offers the possibility of
setting an audible alarm delay. You can specify a short waiting
period (0 to 15 s) after an alarm condition occurs before the
system sounds an audible alarm, thereby potentially reducing
the number of audible SpO
2
alarms.
A visual alarm is still generated anytime the parameter is above
or below the limits set; the delay affects whether the system
sounds an audible alarm for the condition.
When enabled, the SpO
2
value must be above or below the
specified alarm limits for the selected amount of time before
the audible alarm sounds.
An SpO2 alarm delay is set in the
System -> SpO
2
window.
See Section F.8.1.4.

Table of Contents

Related product manuals