3-22 624074/07
3 Hardware options
3.4.6 Continuous nebulization
NOTE:
DO NOT remove silicone plug from nebulizer body.
Before use perform a functional test of Aeroneb use as
described here. Connect the nebulizer (Section 3.4.5) before
proceeding.
1. Ensure the nebulizer unit is firmly fitted into the Aeroneb
Solo T-piece in the breathing circuit.
2. Remove the cap from the medication-filled syringe.
3. Attach the syringe end of the tubing onto the syringe.
4. Prime the tubing completely.
5. Screw the nebulizer end of the tubing onto the top of the
nebulizer.
6. Insert the syringe of medication into the infusion pump.
7. Turn on the continuous mode option on the Aeroneb mod-
ule.
8. Observe nebulizer for proper operation. During continuous
nebulization, the nebulizer is on continuously and the
medication is nebulized on a drop by drop basis. Nebuliza-
tion should be visible with regular intermittent pauses.
Medication level in the nebulizer reservoir should not rise
during use.