12-6 624074/07
12 Special functions
1. To perform the suctioning maneuver, press the
O
2
enrichment
key for pre-oxygenation.
Disconnection terminates the ventilation and starts the
maneuver. For 60 seconds, all alarms are completely
suppressed.
After reconnection the post-oxygenation starts and again
for 60 seconds all acoustic alarms are suppressed. The
alarm message and lamp is still active.
2. To terminate the pre- and post oxygenation maneuver,
press the
O
2
enrichment
key.
12.4 Manual breath
The
Manual breath
key lets the device deliver a manually
triggered breath. The manual breath uses the same settings
of a mandatory breath (standard or operator-set). You can
activate this function in all modes of ventilation.
To deliver a manual breath, press and release the key. DO NOT
quickly or repeatedly press the key.
If you try to start a manual breath during the early stage of
inspiration or the early stage of exhalation, the breath will not
be delivered.
12.5 Pneumatic and Aeroneb
®
nebulizer
WARNING
• DO NOT use an expiratory filter or HME in the
patient’s breathing circuit during nebulization.
Nebulization can cause an expiratory side filter
to clog, substantially increasing flow resistance
and impairing ventilation.
• To prevent the expiratory valve from sticking
due to nebulized medications, use only medica-
tions approved for nebulization and regularly
check and clean the expiratory valve.